EU MDR 2017/745 compliance specialist for medical device classification, technical documentation, clinical evidence, and post-market surveillance. Covers Annex VIII classification rules, Annex II/III technical files, Annex XIV clinical evaluation, and EUDAMED integration.

适用职业律师与法务
License 许可
MIT
最近更新
2026-04-01
Stars
★ 3
Forks
2

01 前言 Front Matter

name
mdr-745-specialist
triggers
[ "MDR compliance", "EU MDR", "medical device classification", "Annex VIII", "technical documentation", "clinical evaluation", "PMCF", "EUDAMED", "UDI", "notified body" ]
description
EU MDR 2017/745 compliance specialist for medical device classification, technical documentation, clinical evidence, and post-market surveillance. Covers Annex VIII classification rules, Annex II/III technical files, Annex XIV clinical evaluation, and EUDAMED integration.

02 资源文件 4

03 SKILL.md 正文

MDR 2017/745 Specialist

EU MDR compliance patterns for medical device classification, technical documentation, and clinical evidence.


Table of Contents


Device Classification Workflow

Classify device under MDR Annex VIII:

  1. Identify device duration (transient, short-term, long-term)
  2. Determine invasiveness level (non-invasive, body orifice, surgical)
  3. Assess body system contact (CNS, cardiac, other)
  4. Check if active device (energy dependent)
  5. Apply classification rules 1-22
  6. For software, apply MDCG 2019-11 algorithm
  7. Document classification rationale
  8. Validation: Classification confirmed with Notified Body

Classification Matrix

Factor Class I Class IIa Class IIb Class III
Duration Any Short-term Long-term Long-term
Invasiveness Non-invasive Body orifice Surgical Implantable
System Any Non-critical Critical organs CNS/cardiac
Risk Lowest Low-medium Medium-high Highest

Software Classification (MDCG 2019-11)

Information Use Condition Severity Class
Informs decision Non-serious IIa
Informs decision Serious IIb
Drives/treats Critical III

Classification Examples

Example 1: Absorbable Surgical Suture

  • Rule 8 (implantable, long-term)
  • Duration: > 30 days (absorbed)
  • Contact: General tissue
  • Classification: Class IIb

Example 2: AI Diagnostic Software

  • Rule 11 + MDCG 2019-11
  • Function: Diagnoses serious condition
  • Classification: Class IIb

Example 3: Cardiac Pacemaker

  • Rule 8 (implantable)
  • Contact: Central circulatory system
  • Classification: Class III

Technical Documentation

Prepare technical file per Annex II and III:

  1. Create device description (variants, accessories, intended purpose)
  2. Develop labeling (Article 13 requirements, IFU)
  3. Document design and manufacturing process
  4. Complete GSPR compliance matrix
  5. Prepare benefit-risk analysis
  6. Compile verification and validation evidence
  7. Integrate risk management file (ISO 14971)
  8. Validation: Technical file reviewed for completeness

Technical File Structure

ANNEX II TECHNICAL DOCUMENTATION
├── Device description and UDI-DI
├── Label and instructions for use
├── Design and manufacturing info
├── GSPR compliance matrix
├── Benefit-risk analysis
├── Verification and validation
└── Clinical evaluation report

GSPR Compliance Checklist

Requirement Evidence Status
Safe design (GSPR 1-3) Risk management file ☐
Chemical properties (GSPR 10.1) Biocompatibility report ☐
Infection risk (GSPR 10.2) Sterilization validation ☐
Software requirements (GSPR 17) IEC 62304 documentation ☐
Labeling (GSPR 23) Label artwork, IFU ☐

Conformity Assessment Routes

Class Route NB Involvement
I Annex II self-declaration None
Is/Im Annex II + IX/XI Sterile/measuring aspects
IIa Annex II + IX or XI Product or QMS
IIb Annex IX + X or X + XI Type exam + production
III Annex IX + X Full QMS + type exam

Clinical Evidence

Develop clinical evidence strategy per Annex XIV:

  1. Define clinical claims and endpoints
  2. Conduct systematic literature search
  3. Appraise clinical data quality
  4. Assess equivalence (technical, biological, clinical)
  5. Identify evidence gaps
  6. Determine if clinical investigation required
  7. Prepare Clinical Evaluation Report (CER)
  8. Validation: CER reviewed by qualified evaluator

Evidence Requirements by Class

Class Minimum Evidence Investigation
I Risk-benefit analysis Not typically required
IIa Literature + post-market May be required
IIb Systematic literature review Often required
III Comprehensive clinical data Required (Article 61)

Clinical Evaluation Report Structure

CER CONTENTS
├── Executive summary
├── Device scope and intended purpose
├── Clinical background (state of the art)
├── Literature search methodology
├── Data appraisal and analysis
├── Safety and performance conclusions
├── Benefit-risk determination
└── PMCF plan summary

Qualified Evaluator Requirements

  • Medical degree or equivalent healthcare qualification
  • 4+ years clinical experience in relevant field
  • Training in clinical evaluation methodology
  • Understanding of MDR requirements

Post-Market Surveillance

Establish PMS system per Chapter VII:

  1. Develop PMS plan (Article 84)
  2. Define data collection methods
  3. Establish complaint handling procedures
  4. Create vigilance reporting process
  5. Plan Periodic Safety Update Reports (PSUR)
  6. Integrate with PMCF activities
  7. Define trend analysis and signal detection
  8. Validation: PMS system audited annually

PMS System Components

Component Requirement Frequency
PMS Plan Article 84 Maintain current
PSUR Class IIa and higher Per class schedule
PMCF Plan Annex XIV Part B Update with CER
PMCF Report Annex XIV Part B Annual (Class III)
Vigilance Articles 87-92 As events occur

PSUR Schedule

Class Frequency
Class III Annual
Class IIb implantable Annual
Class IIb Every 2 years
Class IIa When necessary

Serious Incident Reporting

Timeline Requirement
2 days Serious public health threat
10 days Death or serious deterioration
15 days Other serious incidents

EUDAMED and UDI

Implement UDI system per Article 27:

  1. Obtain issuing entity code (GS1, HIBCC, ICCBBA)
  2. Assign UDI-DI to each device variant
  3. Assign UDI-PI (production identifier)
  4. Apply UDI carrier to labels (AIDC + HRI)
  5. Register actor in EUDAMED
  6. Register devices in EUDAMED
  7. Upload certificates when available
  8. Validation: UDI verified on sample labels

EUDAMED Modules

Module Content Actor
Actor Company registration Manufacturer, AR
UDI/Device Device and variant data Manufacturer
Certificates NB certificates Notified Body
Clinical Investigation Study registration Sponsor
Vigilance Incident reports Manufacturer
Market Surveillance Authority actions Competent Authority

UDI Label Requirements

Required elements per Article 13:

  • [ ] UDI-DI (device identifier)
  • [ ] UDI-PI (production identifier) for Class II+
  • [ ] AIDC format (barcode/RFID)
  • [ ] HRI format (human-readable)
  • [ ] Manufacturer name and address
  • [ ] Lot/serial number
  • [ ] Expiration date (if applicable)

Reference Documentation

MDR Classification Guide

references/mdr-classification-guide.md contains:

  • Complete Annex VIII classification rules (Rules 1-22)
  • Software classification per MDCG 2019-11
  • Worked classification examples
  • Conformity assessment route selection

Clinical Evidence Requirements

references/clinical-evidence-requirements.md contains:

  • Clinical evidence framework and hierarchy
  • Literature search methodology
  • Clinical Evaluation Report structure
  • PMCF plan and evaluation report guidance

Technical Documentation Templates

references/technical-documentation-templates.md contains:

  • Annex II and III content requirements
  • Design History File structure
  • GSPR compliance matrix template
  • Declaration of Conformity template
  • Notified Body submission checklist

Tools

MDR Gap Analyzer

# Quick gap analysis
python scripts/mdr_gap_analyzer.py --device "Device Name" --class IIa

# JSON output for integration
python scripts/mdr_gap_analyzer.py --device "Device Name" --class III --output json

# Interactive assessment
python scripts/mdr_gap_analyzer.py --interactive

Analyzes device against MDR requirements, identifies compliance gaps, generates prioritized recommendations.

Output includes:

  • Requirements checklist by category
  • Gap identification with priorities
  • Critical gap highlighting
  • Compliance roadmap recommendations

Notified Body Interface

Selection Criteria

Factor Considerations
Designation scope Covers your device type
Capacity Timeline for initial audit
Geographic reach Markets you need to access
Technical expertise Experience with your technology
Fee structure Transparency, predictability

Pre-Submission Checklist

  • [ ] Technical documentation complete
  • [ ] GSPR matrix fully addressed
  • [ ] Risk management file current
  • [ ] Clinical evaluation report complete
  • [ ] QMS (ISO 13485) certified
  • [ ] Labeling and IFU finalized
  • [ ] Validation: Internal gap assessment complete

04 同仓库更多 Skills

afaizalam2003
afaizalam2003/Master-claude-skill

soc2-compliance

Use when the user asks to prepare for SOC 2 audits, map Trust Service Criteria, build control matrices, collect audit evidence, perform gap analysis, or assess SOC 2 Type I vs Type II readiness.

律师与法务
★ 3
afaizalam2003
afaizalam2003/Master-claude-skill

risk-management-specialist

Medical device risk management specialist implementing ISO 14971 throughout product lifecycle. Provides risk analysis, risk evaluation, risk control, and post-production information analysis. Use when user mentions risk management, ISO 14971, risk analysis, FMEA, fault tree analysis, hazard identification, risk control, risk matrix, benefit-risk analysis, residual risk, risk acceptability, or post-market risk.

★ 3
afaizalam2003
afaizalam2003/Master-claude-skill

regulatory-affairs-head

Senior Regulatory Affairs Manager for HealthTech and MedTech companies. Prepares FDA 510(k), De Novo, and PMA submission packages; analyzes regulatory pathways for new medical devices; drafts responses to FDA deficiency letters and Notified Body queries; develops CE marking technical documentation under EU MDR 2017/745; coordinates multi-market approval strategies across FDA, EU, Health Canada, PMDA, and NMPA; and maintains regulatory intelligence on evolving standards. Use when users need to plan or execute FDA submissions, navigate 510(k) or PMA approval processes, achieve CE marking, prepare pre-submission meeting materials, write regulatory strategy documents, respond to agency queries, or manage compliance documentation for medical device market access.

★ 3
afaizalam2003
afaizalam2003/Master-claude-skill

quality-manager-qms-iso13485

ISO 13485 Quality Management System implementation and maintenance for medical device organizations. Provides QMS design, documentation control, internal auditing, CAPA management, and certification support. Use when working with medical device quality systems, preparing for ISO 13485 audits, managing regulatory compliance documentation, setting up corrective actions, or building audit preparation programs. Useful for quality management, audit preparation, regulatory compliance, medical device documentation, and corrective action workflows.

设计师视觉设计
★ 3
afaizalam2003
afaizalam2003/Master-claude-skill

quality-manager-qmr

Senior Quality Manager Responsible Person (QMR) for HealthTech and MedTech companies. Provides quality system governance, management review leadership, regulatory compliance oversight, and quality performance monitoring per ISO 13485 Clause 5.5.2.

律师与法务
★ 3
afaizalam2003
afaizalam2003/Master-claude-skill

quality-documentation-manager

Document control system management for medical device QMS. Covers document numbering, version control, change management, and 21 CFR Part 11 compliance. Use for document control procedures, change control workflow, document numbering, version management, electronic signature compliance, or regulatory documentation review.

律师与法务文档处理Skill与智能体开发
★ 3